CASP-50 - 产品中心 - 北京凯斯普医疗器械有限公司,牡丹江等离子体物理应用科技有限公司

CASP-50

Content

Name: Hydrogen Peroxide Low Temperature Plasma Sterilizer

Model: CASP-50
Electronic Requirements:AC220/AC230V, 50/60HZ

Power Consumption: 3.5KW

Monitor: 7’’ TFT Color LCD Touch Screen

Overall Size(mm): L850 × W895 × H1660
Weight: 360kg

Chamber Materizal: Aluminum Alloy

Chamber Size(mm): L680 × W396 × H236

Chamber Volume: 64L

Usable Chamber Volume: 55L
Shelf Size(mm): L633 × W343 × 2EA(Movable)

Sterilization Agent: 55~60% Hydrogen Peroxide

Injection per Cycle:
Fast: 2.5ml×1time

Normal/Super: 2.5ml×2times

Cycle Temperature Range: 55±5

Sterilization Cycle Time:

Fast: 18min

⊙Instruments with stainless steel lumen(two-side opened), I.D. ≥0.7mm, L ≤200mm
⊙Instruments with polyethylene or Teflon lumen(two-side opened), I.D. ≥1mm, L ≤ 2m


Normal: 42min

⊙Instruments with stainless steel lumen(two-side opened), I.D. ≥0.7mm, L ≤300mm
⊙Instruments with polyethylene or Teflon lumen (two-side opened), I.D. ≥1mm, L ≤ 3m


Super: 52min

⊙Instruments with stainless steel lumen(two-side opened), I.D.≥0.7mm, L≤ 600mm
⊙Instruments with polyethylene or Teflon lumen (two-side opened), I.D.≥1mm, L≤4m

Data Storage: 60,000 cycles

Noise: 60dB

Sterilization Guarantee:
Stainless Steel Lumen

I.D.≥0.7mm

Length≤60cm

Teflon Lumen

I.D.≥1mm

Length≤400cm